انکوباتورهای ممرت آلمان در ابعاد و حجم های مختلف
IPP 400 انکوباتور
Stand delivery Peltier-Cooled incubator
Our Peltier-cooled incubator offers unparalleled precision and control accuracy and is energy-saving and gentle. The Peltier technology minimises the risk of the sample drying out, condensation formation during the cooling down process takes place outside the working chamber, on the Peltier element.
Ventilation and Contorl
forced ventilation by Peltier fan
adaptive, fuzzy-supported microprocessor PIDcontroller, for stepless control of Peltier-heating-/ cooling system without refrigerant
autodiagnostic system with fault indication
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
digital 7-day-programme-timer with real time clock, precise minute setting, for one set value or start of ramp operation
integrated timer for tempering profiles of up to 40 ramps, each segment adjustable from 1 min. up to 999 hours
multifunctional programming via menu on 8-digit alphanumeric digital display (language to be chosen via set-up):
tempering profiles of up to 40 ramps
time- and set-point dependent operation
digital display (LED) of all set parameters, such as temperature, weekdays, time, programme status and set-up values
resolution of display for set value and actual value 0,1 C
long-term logging (ring store) of all relevant data, GLP-conforming as data logger 1024 kB
programme stored on power failure
parallel printer interface (incl. real-time clock with date function) for printing logging files, suitable all PCL3-compatible ink-jet printers (USB available via converter, see accessories)
USB interface including MEMMERT Software Celsius for programming and documentation
chip card control incl. one MEMoryCard XL with 32 kB (up to 40 ramps)
incl. works calibration certificate for +10 C and +37 C
Multiple Overtemperature Protection
with audible and visual alarm
independently working, digitally adjustable electronic microprocessor overtemperature controller TWW protection class 3.3 maximum value for overtemperature, minimum value for undertemperature
additional integral over- and undertemperature protection ASF (Auto-Safety-Function) automatically following the set value at a preset tolerance range; audible alarm is activated in case of over- or undertemperature, heating function is switched off in case of overtemperature cooling function is switched off in case of undertemperature
resolution of display and setting accuracy: 0,1 C
Textured Staninless Steel Casing
w x h x d: 550 x 680 x 570 mm (+38 mm for handle)
fully insulated stainless steel door with double locking and 4-point adjustment
inner glass door
rear zinc-plated steel
Interior-Heating Concept
w x h x d: 550 x 680 x 570 mm (+38 mm for handle)
fully insulated stainless steel door with double locking and 4-point adjustment
inner glass door
rear zinc-plated steel
Temperature Range
from +5 C up to +70 C
temperature variation in time: < +/-0,1 C (to DIN 12 880: 2007-05)
temperature uniformity in chamber at 10C and 37C: < +/- 0.4 C (to DIN 12 880: 2007-05)
Voltage/Power Rating
230 V (+/- 10%), 50/60 Hz
ca. 175 W (during heating and cooling)
Packing Data
net weight approx. 43 kg
gross weight carton approx. 49 kg
dimensions approx.:
carton w x h x d: 67 x 78 x 68 cm
the appliances must be transported upright
Customs Tariff Number
8419 8998
Country of Origin
Federal Republic of Germany
WEEE-Reg.NO.
DE 66812464
Accessories
Perforated stainless steel shelf E0(x)
Stainless Steel Grid E3(x)
RS232 interface instead of USB W6
Interface Ethernet instead of USB inclusive software Celsius Ethernet-Edition W4
Parallel/USB converter cable with integrated power supply unit to connect PCL3-compatible HP printers with USB interface to MEMMERT units. W1
Documentation package consisting of parallel USB converter cable including PLC3-compatible HP colour inkjet printer with USB interface (HP Deskjet 6980 or successor) for direct connection of printer to Memmert unit W2
IQ check list with works test data for oven as support for validation by customer Q1
OQ check list including one free-selectable temperature distribution survey to DIN 12880: 2007-05 (size 200/300: 9 measuring points, size 400-800: 27 measuring points) with works test data for oven as support for validation by customer Q2
Software conforming to FDA Celsius FDA-Edition for up to 16 units Q3
Oven-linked authorisation card (User-ID-Card) prevents undesired manipulation by unauthorised third parties V1
Computer interface RS485 (for networking a maximum of 16 ovens) instead of interface USB V2
Temperature profile write/read unit for programming via PC, for writing to and reading from the chip card, up to 40 ramps V3
Additional chip card, blank, formatted (32 kB MEMoryCard XL for a maximum of 40 ramps) V4